7
Specifications
184
Pulse Laser Station NPL-302 Series Instruction Manual
To connect to an external power source, supply power to Pin 3 (power terminal) and
Pin 5 (ground terminal) on the instrument. The instrument will use the external
power source even if the carrying handle battery BC-65 is attached.
C
Caution –
Make sure that the power supplied is within the rated input range
(7.2 V to 11 V, 1 A maximum). Power supplied outside this range will damage
the instrument.
To communicate with an external device, connect the RS-232C signal from the
external device to Pin 1 (input terminal) and to Pin 2 (output terminal) on the
instrument.
Cap the data output/external power input connector securely while not in use. The
instrument is not watertight if the cap is not attached or not attached securely, and
when the data output/external power input connector is in use.
The instrument can be damaged by static electricity from the human body
discharged through the data output/external power input connector. Before handling
the instrument, touch any other conductive material once to remove static
electricity.
Содержание NPL-302 Series
Страница 1: ...H163E 05 10 TF 9 Pulse Laser Station NPL 302 Series Instruction Manual NPL 332 NPL 352 NPL 362 ...
Страница 4: ......
Страница 5: ...Safety 1 In this chapter Q Introduction Q Laser Safety Q Warnings and Cautions ...
Страница 12: ...Safety xii Pulse Laser Station NPL 302 Series Instruction Manual ...
Страница 28: ...1 Introduction 8 Pulse Laser Station NPL 302 Series Instruction Manual ...
Страница 46: ...2 Preparation 26 Pulse Laser Station NPL 302 Series Instruction Manual ...
Страница 82: ...3 Getting Started 62 Pulse Laser Station NPL 302 Series Instruction Manual ...
Страница 188: ...5 Menu Key 168 Pulse Laser Station NPL 302 Series Instruction Manual ...
Страница 196: ...6 Checking and Adjustment 176 Pulse Laser Station NPL 302 Series Instruction Manual ...
Страница 208: ...8 System Diagrams 188 Pulse Laser Station NPL 302 Series Instruction Manual ...