APPENDIX
345
Measuring Characteristics
DC Voltage
Input Resistance Range
100 mV
10 M
Ω
or >10 G
Ω
Selectable
1 V
10 V
100 V
10
MΩ±1%
1000 V
Input Bias
30 pA (Typ, 25°C)
Input Protection
1000 V on all ranges
Measurement Method: Sigma-delta A/D Converter
Resistance
Max. Lead
Resistance
10% of range per lead for 100
Ω
, 1 k
Ω
ranges. 1 k
Ω
per lead on all other ranges.
Input Protection
1000 V on all ranges
Measurement Method: Selectable 4-wire or 2-wire ohms. Current source referenced to Input LO
Terminal
DC Current
Range
Shunt
Burden Voltage
100 μA
100
Ω
<0.011 V
1 mA
100
Ω
<0.11 V
10 mA
1
Ω
<0.04 V
100 mA
1
Ω
<0.4 V
1 A
0.1
Ω
<0.7 V
3 A
0.1
Ω
<2 V
10 A
10m
Ω
<0.5 V
Input Protection
External 3.15 A, 500 V fuse for 3 A
Internal 6 A, 1 kV fuse for 3 A
Internal 12 A, 1 kV fuse for 10 A
Reading Rate
(Readings/sec)
Speed
Digits
DCV
DCI
2W/4W-Resistance
5/s
,
20/s
,
60(50)/s
,
100/s 6 ½
400/s
,
1.2k/s
,
2.4k/s
5 ½
4.8k/s
,
7.2k/s
,
10k/s
4 ¼
Speed
Digits
Continuity
Diode
60(50) /s
6 ½
100 /s
5 ½
400 /s
4 ¼
[1].
DC Specification: In addition to the availability that requires
warm-up of 60 minutes, it must be set in 5/s speed rate (60(50)/s speed
rate for Continuity and Diode), A-Zero on.
[2].
The entire range of measurement will pass the set range by 20%
except the tests of 1000 V DC, 3 A DC, 10 A DC and diode.
[3].
This specification applies to 4-wire resistance measurement, whilst it
requires using "REL" function for offset on 2-wire resistance
measurement. 2-wire resistance measurement will cause additional error
Содержание DM2571
Страница 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Страница 2: ......
Страница 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 76: ...DM2571 User Manual 72 ...
Страница 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Страница 176: ...DM2571 User Manual 172 ...
Страница 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Страница 185: ...DISPLAY SETTING 181 DISPLAY SETTING Digit 182 Display 184 Number 184 Bar Meter 185 Trend Chart 189 Histogram 198 ...
Страница 244: ...DM2571 User Manual 240 ...
Страница 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Страница 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Страница 348: ...DM2571 User Manual 344 DCV Ratio 5 Accuracy Specification DC Input accuracy DC Reference accuracy ...
Страница 360: ...DM2571 User Manual 356 Dimensions All dimensions are shown in millimeters Optional GPIB card is installed ...
Страница 362: ......
Страница 363: ......
Страница 364: ......
Страница 366: ......
Страница 368: ......
Страница 369: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama223 8508JAPAN ...