DM2571 User Manual
304
SENSe TEMPerature Commands
[SENSe:]TEMPerature:NPLCycles
Sets the integration time for Temperature measurement in PLCs (power line
cycles). Where one PLC is equal to 16.6 milliseconds. For any <NRf>
parameter, the DMM will automatically set the PLC to the closest acceptable
PLC value (1 | 3 | 12).
Parameter: NPLCycles(<NRf> | MIN | MAX | DEF)
Example: SENS:TEMP:NPLC DEF
Sets the integration time to the 12 PLCs for Temperature measurement.
[SENSe:]TEMPerature:NPLCycles?
Returns the integration time for Temperature measurement in PLCs (power
line cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 1 | 3 | 12
[SENSe:]TEMPerature:NULL[:STATe]
Sets the relative function on/off for Temperature measurement.
Parameter: 0 | 1 | ON | OFF
Example: SENS:TEMP:NULL:STAT ON
Turns the relative function on for Period measurement.
[SENSe:]TEMPerature:NULL[:STATe]?
Returns the relative function state of Temperature measurement.
Return parameter: 0 | 1, 0=OFF, 1=ON
[SENSe:]TEMPerature:NULL:VALue
Sets the relative value for Temperature measurement.
Parameter: <NRf> (-1.0e15 to 1.0e15) | MIN | MAX | DEF
Example: SENS:FREQ:NULL:VAL 5
Sets the relative value to 5
˚
C for Temperature measurement.
[SENSe:]TEMPerature:NULL:VALue?
Returns the current relative value of Temperature measurement.
[SENSe:]TEMPerature:NULL:VALue:AUTO
Sets the relative value auto on/off for Temperature measurement.
Parameter: 0 | 1 | ON | OFF
Example: SENS:TEMP:NULL:STAT ON
SENS:TEMP:NULL:VAL:AUTO ON
The unit automatically sets the 1st count of measurement as null value.
[SENSe:]TEMPerature:NULL:VALue:AUTO?
Returns the null value auto state of Temperature measurement.
Содержание DM2571
Страница 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Страница 2: ......
Страница 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 76: ...DM2571 User Manual 72 ...
Страница 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Страница 176: ...DM2571 User Manual 172 ...
Страница 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Страница 185: ...DISPLAY SETTING 181 DISPLAY SETTING Digit 182 Display 184 Number 184 Bar Meter 185 Trend Chart 189 Histogram 198 ...
Страница 244: ...DM2571 User Manual 240 ...
Страница 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Страница 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Страница 348: ...DM2571 User Manual 344 DCV Ratio 5 Accuracy Specification DC Input accuracy DC Reference accuracy ...
Страница 360: ...DM2571 User Manual 356 Dimensions All dimensions are shown in millimeters Optional GPIB card is installed ...
Страница 362: ......
Страница 363: ......
Страница 364: ......
Страница 366: ......
Страница 368: ......
Страница 369: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama223 8508JAPAN ...