REMOTE CONTROL
269
CONFigure[:VOLTage][:DC]:RATio
Sets measurement to DCV ratio mode on the 1
st
display and specifies
range/resolution.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX |
DEF)[,Resolution(<NRf> | MIN | MAX | DEF)]]
Example: CONF:VOLT:DC:RAT 1
Sets the DC voltage range to 1V using the default resolution.
●Autoranging (AUTO or DEFault), will generate an error if you specify a
<resolution> because the instrument cannot accurately resolve the integration
time (especially if the input continuously changes). If your application
requires autoranging, specify DEFault for the <resolution> or omit the
<resolution> altogether.
CONFigure[:VOLTage]:AC
Sets measurement to AC Voltage on the 1
st
display and specifies range.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF)]
Example: CONF:VOLT:AC
Sets the AC voltage range to auto range.
CONFigure:CURRent[:DC]
Sets measurement to DC Current on the 1
st
display and specifies
range/resolution.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX |
DEF)[,Resolution(<NRf> | MIN | MAX | DEF)]]
Example: CONF:CURR:DC 10e-3,DEF
Sets the DC current range to 10mA using the default resolution.
●Autoranging (AUTO or DEFault), will generate an error if you specify a
<resolution> because the instrument cannot accurately resolve the integration
time (especially if the input continuously changes). If your application
requires autoranging, specify DEFault for the <resolution> or omit the
<resolution> altogether.
CONFigure:CURRent:AC
Sets measurement to AC Current on the 1
st
display and specifies range.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF)]
Example: CONF:CURR:AC 10e-2
Sets the measurement mode to AC Current with a 100mA range.
Содержание DM2571
Страница 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Страница 2: ......
Страница 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 76: ...DM2571 User Manual 72 ...
Страница 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Страница 176: ...DM2571 User Manual 172 ...
Страница 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Страница 185: ...DISPLAY SETTING 181 DISPLAY SETTING Digit 182 Display 184 Number 184 Bar Meter 185 Trend Chart 189 Histogram 198 ...
Страница 244: ...DM2571 User Manual 240 ...
Страница 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Страница 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Страница 348: ...DM2571 User Manual 344 DCV Ratio 5 Accuracy Specification DC Input accuracy DC Reference accuracy ...
Страница 360: ...DM2571 User Manual 356 Dimensions All dimensions are shown in millimeters Optional GPIB card is installed ...
Страница 362: ......
Страница 363: ......
Страница 364: ......
Страница 366: ......
Страница 368: ......
Страница 369: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama223 8508JAPAN ...