APPENDIX
353
Measuring Characteristics
Speed
Digits
Reading Rate
(Readings/sec)
RTD/TCO/
Thermistor
5/s
6 ½
20/s
5 ½
60(50)/s
4 ½
[1].
The actual measurement range and test lead error will be constrained
by the adopted test lead. The test lead accuracy adder covers all errors of
measurements and ITS-90 temerature change.
Capacitance
[1]
Range
1 Year
TCAL ± 5 ˚C
Temperature
Coefficient/
˚
C
1.000 nF
[2]
2.00 + 2.00
0.05 + 0.01
10.00 nF
2.00 + 1.00
0.05 + 0.01
100.0 nF
2.00 + 0.40
0.05 + 0.01
1.000 μF
2.00 + 0.40
0.05 + 0.01
10.00 μF
2.00 + 0.40
0.05 + 0.01
100.0 μF
2.00 + 0.40
0.05 + 0.01
Accuracy Specifications: ± ( % of r % of range )
[1].
Specifications are for film capacitor inputs that are greater than 10% range.
[2].
Accuracy at 1nF range is supplementary value.
Measuring Characteristics
Measurement method: DC recharge & discharge.
Input protection: 500 Vpeak on all ranges.
The capacitor under test (Cx) is charged using a constant current source. The
time to charge Cx is recorded. The capacitor is then discharged using a known
resistance and the discharge time is recorded. The value of the resistance depends
on the capacitance range that is selected. The charge and discharge time is used
to calculate the capacitance of Cx if the selected capacitance range is equal to or
less than 10 nF. Only the charge time is used to calculate the capacitance of Cx if
the selected capacitance range is equal to or greater than 100 nF.
As measuring capacitance with the DMM is effectively a DC measurement, the
measured capacitance tends to be higher than what is measured by LCR meters.
For best measurement results, first perform a zeroing of the test leads when the
cables are “open” to compensate for the test lead capacitance.
Содержание DM2571
Страница 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Страница 2: ......
Страница 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 76: ...DM2571 User Manual 72 ...
Страница 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Страница 176: ...DM2571 User Manual 172 ...
Страница 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Страница 185: ...DISPLAY SETTING 181 DISPLAY SETTING Digit 182 Display 184 Number 184 Bar Meter 185 Trend Chart 189 Histogram 198 ...
Страница 244: ...DM2571 User Manual 240 ...
Страница 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Страница 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Страница 348: ...DM2571 User Manual 344 DCV Ratio 5 Accuracy Specification DC Input accuracy DC Reference accuracy ...
Страница 360: ...DM2571 User Manual 356 Dimensions All dimensions are shown in millimeters Optional GPIB card is installed ...
Страница 362: ......
Страница 363: ......
Страница 364: ......
Страница 366: ......
Страница 368: ......
Страница 369: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama223 8508JAPAN ...