[SENSe:]VOLTage[:DC]:NULL:VALue
Sets the relative value for DC Voltage measurement.
Parameter: <NRf> (-1200.0 to 1200.0 V) | MIN | MAX | DEF
Example: SENS:VOLT:DC:NULL:STAT ON
SENS:VOLT:DC:NULL:VAL 1.2
Sets the relative value to 1.2V for DC Voltage measurement.
[SENSe:]VOLTage[:DC]:NULL:VALue?
Returns the current relative value of DC Voltage measurement.
[SENSe:]VOLTage[:DC]:NULL:VALue:AUTO
Sets the relative value auto on/off for DC Voltage measurement.
Parameter: 0 | 1 | ON | OFF
Example: SENS:VOLT:DC:NULL:STAT ON
SENS:VOLT:DC:NULL:VAL:AUTO ON
READ ?
The unit automatically sets the 1st count of measurement as null value.
[SENSe:]VOLTage[:DC]:NULL:VALue:AUTO?
Returns the null value auto state of DC Voltage measurement.
[SENSe:]VOLTage[:DC]:RANGe
Sets the DC voltage measurement range.
Parameter: <NRf> | MIN | MAX | DEF
Example: SENS:VOLT:DC:RANG MIN
Set the DC voltage range to lowest range allowed.
[SENSe:]VOLTage[:DC]:RANGe?
Returns the DC voltage measurement range.
[SENSe:]VOLTage[:DC]:RANGe:AUTO
Sets the DC voltage Auto-range setting on, off or once only.
Parameter: 0 | 1 | ON | OFF | ONCE
Example: SENS:VOLT:DC:RANG:AUTO ON
Turns Auto-range on for DC voltage measurements.
[SENSe:]VOLTage[:DC:]RANGe:AUTO?
Returns the DC voltage Auto-range settings.
Return parameter: 0 | 1, 0=OFF, 1=ON
[SENSe:]VOLTage[:DC]:RESolution
Sets the DC Voltage measurement resolution. The resolution depends on the
rate and range settings.
Parameter: Resolution(<NRf> | MIN | MAX | DEF)
Example: SENS:VOLT:DC:RES MAX
Sets the DC Voltage resolution to MAX.
[SENSe:]VOLTage[:DC]:RESolution?
Returns the DC Voltage resolution.
Содержание DM2571
Страница 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Страница 2: ......
Страница 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 76: ...DM2571 User Manual 72 ...
Страница 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Страница 176: ...DM2571 User Manual 172 ...
Страница 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Страница 185: ...DISPLAY SETTING 181 DISPLAY SETTING Digit 182 Display 184 Number 184 Bar Meter 185 Trend Chart 189 Histogram 198 ...
Страница 244: ...DM2571 User Manual 240 ...
Страница 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Страница 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Страница 348: ...DM2571 User Manual 344 DCV Ratio 5 Accuracy Specification DC Input accuracy DC Reference accuracy ...
Страница 360: ...DM2571 User Manual 356 Dimensions All dimensions are shown in millimeters Optional GPIB card is installed ...
Страница 362: ......
Страница 363: ......
Страница 364: ......
Страница 366: ......
Страница 368: ......
Страница 369: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama223 8508JAPAN ...