DM2571 User Manual
292
[SENSe:]CURRent[:DC]:RANGe
Sets the DC current measurement range.
Parameter: Range(<NRf> | MIN | MAX | DEF)
Example: SENS:CURR:DC:RANG 10e-2
Sets the DC current range to 100mA.
[SENSe:]CURRent[:DC]:RANGe?
Returns the DC current measurement range.
[SENSe:]CURRent[:DC]:RANGe:AUTO
Sets the DC current Auto-range settings on, off or once.
Parameter: 0 | 1 | ON | OFF | ONCE
Example: SENS:CURR:DC:RANG:AUTO OFF
Turns Auto-range off for DC current measurement.
[SENSe:]CURRent[:DC]:RANGe:AUTO?
Returns the DC current Auto-range settings.
Return parameter: 0 | 1, 0=OFF, 1=ON
[SENSe:]CURRent[:DC]:RESolution
Sets the DC Current measurement resolution. The resolution depends on the
rate and range settings.
Parameter: Resolution(<NRf> | MIN | MAX | DEF)
Example: SENS:CURR:DC:RES 0.01
Sets the DC Current resolution to 0.01
[SENSe:]CURRent[:DC]:RESolution?
Returns the DC Current resolution.
[SENSe:]CURRent[:DC]:TERMinals
Assigns an input port for the current function.
Parameter: <NR1> 3 | 10
Example: SENS:CURR:DC:TERM 3
Sets the input jack to the 3A current input port.
[SENSe:]CURRent[:DC]:TERMinals?
Returns the assigned input port used for the current function.
Return parameter: +3 | +10
[SENSe:]CURRent[:DC]:TRIGger:DELay
Sets the trigger delay time that minimum step is microseconds of DC Current
measurement.
Parameter: <NRf>(0 to 3600 s) | MIN | MAX | DEF
Example: SENS:CURR:DC:TRIG:DEL 2e-4
Sets the trigger delay time to 200us of DC Current measurement.
[SENSe:]CURRent[:DC]:TRIGger:DELay?
Returns the trigger delay time in seconds of DC Current measurement.
Return parameter: <NRf>
Содержание DM2571
Страница 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Страница 2: ......
Страница 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 76: ...DM2571 User Manual 72 ...
Страница 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Страница 176: ...DM2571 User Manual 172 ...
Страница 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Страница 185: ...DISPLAY SETTING 181 DISPLAY SETTING Digit 182 Display 184 Number 184 Bar Meter 185 Trend Chart 189 Histogram 198 ...
Страница 244: ...DM2571 User Manual 240 ...
Страница 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Страница 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Страница 348: ...DM2571 User Manual 344 DCV Ratio 5 Accuracy Specification DC Input accuracy DC Reference accuracy ...
Страница 360: ...DM2571 User Manual 356 Dimensions All dimensions are shown in millimeters Optional GPIB card is installed ...
Страница 362: ......
Страница 363: ......
Страница 364: ......
Страница 366: ......
Страница 368: ......
Страница 369: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama223 8508JAPAN ...