[SENSe:]CONTinuity:ZERO:AUTO
Sets the auto zero mode to on, off or once only of Continuity measurement.
Parameter: 0 | 1 | ON | OFF | ONCE
Example: SENS:CONT:ZERO:AUTO OFF
Sets the auto zero to off.
[SENSe:]CONTinuity:ZERO:AUTO?
Returns the auto zero mode of Continuity measurement.
Return parameter: 0 | 1, 1=ON, 0=OFF
SENSe DIODe Commands
[SENSe:]DIODe:NPLCycles
Sets the integration time for Diode measurement in PLCs (power line cycles).
Where one PLC is equal to 16.6 milliseconds. For any <NRf> parameter, the
DMM will automatically set the PLC to the closest acceptable PLC value
(0.15 | 0.6 | 1).
Parameter: NPLCycles(<NRf> | MIN | MAX | DEF)
Example: SENS:DIOD:NPLC DEF
Sets the integration time to the 1 PLCs for diode measurement.
[SENSe:]DIODe:NPLCycles?
Returns the integration time for Diode measurement in PLCs (power line
cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 0.15 | 0.6 | 1
[SENSe:]DIODe:RESolution
Sets the Diode measurement resolution. The resolution depends on the rate
and range settings.
Parameter: Resolution(<NRf> | MIN | MAX | DEF)
Example:SENS:DIOD:RES 0.1e-4
Sets the Diode resolution to 0.00001
[SENSe:]DIODe:RESolution?
Returns the Diode measurement resolution.
[SENSe:]DIODe:TRIGger:DELay
Sets the trigger delay that minimum step is microseconds of Diode
measurement.
Parameter: <NRf>(0 to 3600 s) | MIN | MAX | DEF
Example: SENS:DIOD:TRIG:DEL 0.5
Sets the trigger delay time to 500ms of Diode measurement.
[SENSe:]DIODe:TRIGger:DELay?
Returns the trigger delay time in seconds of Diode measurement.
Return parameter: <NRf>
Содержание DM2571
Страница 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Страница 2: ......
Страница 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 76: ...DM2571 User Manual 72 ...
Страница 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Страница 176: ...DM2571 User Manual 172 ...
Страница 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Страница 185: ...DISPLAY SETTING 181 DISPLAY SETTING Digit 182 Display 184 Number 184 Bar Meter 185 Trend Chart 189 Histogram 198 ...
Страница 244: ...DM2571 User Manual 240 ...
Страница 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Страница 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Страница 348: ...DM2571 User Manual 344 DCV Ratio 5 Accuracy Specification DC Input accuracy DC Reference accuracy ...
Страница 360: ...DM2571 User Manual 356 Dimensions All dimensions are shown in millimeters Optional GPIB card is installed ...
Страница 362: ......
Страница 363: ......
Страница 364: ......
Страница 366: ......
Страница 368: ......
Страница 369: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama223 8508JAPAN ...