DM2571 User Manual
274
DATA:REMove? <reading_number>,[WAIT]
Reads and erases measurement values from reading memory up to the
specified <reading_number>.
The measurement values are read and erased from the reading memory starting
with the oldest measurement first.
Ex:SAMP:COUN 10
INIT
DATA :REM? 4
Returns:
-1.12816521E-04,-1.13148354E-04,-1.13485152E-04,-1.13365632E-04
●If you do not specify a value for <reading_number>, +9.9137 (Not
a Number) is returned.
●If reading_number is greater than the latest counts of measurement, it will
return the error. However, it will return data if reading_number of counts of
measurement reach the set threshold only when WAIT parameter is specified.
●The R? and DATA:REMove? queries can be used during a long series of
readings to periodically remove readings from memory that would normally
cause the reading memory to overflow. R? does not wait for all readings to
complete. It sends the readings that are complete at the time the instrument
receives the command.
Содержание DM2571
Страница 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Страница 2: ......
Страница 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 76: ...DM2571 User Manual 72 ...
Страница 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Страница 176: ...DM2571 User Manual 172 ...
Страница 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Страница 185: ...DISPLAY SETTING 181 DISPLAY SETTING Digit 182 Display 184 Number 184 Bar Meter 185 Trend Chart 189 Histogram 198 ...
Страница 244: ...DM2571 User Manual 240 ...
Страница 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Страница 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Страница 348: ...DM2571 User Manual 344 DCV Ratio 5 Accuracy Specification DC Input accuracy DC Reference accuracy ...
Страница 360: ...DM2571 User Manual 356 Dimensions All dimensions are shown in millimeters Optional GPIB card is installed ...
Страница 362: ......
Страница 363: ......
Страница 364: ......
Страница 366: ......
Страница 368: ......
Страница 369: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama223 8508JAPAN ...