DISPlay:TEXT[:DATA]
Sets the text message to TFTLCD display screen.
Parameter: ”<message>”
Example: DISP:TEXT:DATA “testing”
Prints the testing characters to TFT LCD display screen.
DISPlay:TEXT:[:DATA]?
Returns the text message of TFT LCD display screen.
Return parameter: ”<message>”, Ex: ”testing”
DISPlay:VIEW
Sets the display form of measured value.
Parameter: NUMeric | HISTogram | TCHart | METer
Example: DISP:VIEW HIST
Sets display in the histogram mode.
DISPlay:VIEW?
Returns the display form of measured value.
Return parameter: NUM | HIST | TCH | MET
MEASure Commands
MEASure[:VOLTage]:DC?
Returns the DC voltage measurement value on the 1
st
display.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX |
DEF)[,Resolution(<NRf> | MIN | MAX | DEF)]]
Example: MEAS:VOLT:DC? MIN
> +6.64925206E-04
Returns the DC voltage measurement value as 0.6649 mV.
●Autoranging (AUTO or DEFault), will generate an error if you specify a
<resolution> because the instrument cannot accurately resolve the integration
time (especially if the input continuously changes). If your application
requires autoranging, specify DEFault for the <resolution> or omit the
<resolution> altogether.
MEASure[:VOLTage][:DC]:RATio?
Returns the DC ratio measurement value on the 1
st
display.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX |
DEF)[,Resolution(<NRf> | MIN | MAX | DEF)]]
Example: MEAS:VOLT:DC:RAT?
> +2.87393920E-03
Returns the DC ratio measurement value as 2.87393 m.
●Autoranging (AUTO or DEFault), will generate an error if you specify a
<resolution> because the instrument cannot accurately resolve the integration
time (especially if the input continuously changes). If your application
requires autoranging, specify DEFault for the <resolution> or omit the
<resolution> altogether.
Содержание DM2571
Страница 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Страница 2: ......
Страница 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 76: ...DM2571 User Manual 72 ...
Страница 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Страница 176: ...DM2571 User Manual 172 ...
Страница 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Страница 185: ...DISPLAY SETTING 181 DISPLAY SETTING Digit 182 Display 184 Number 184 Bar Meter 185 Trend Chart 189 Histogram 198 ...
Страница 244: ...DM2571 User Manual 240 ...
Страница 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Страница 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Страница 348: ...DM2571 User Manual 344 DCV Ratio 5 Accuracy Specification DC Input accuracy DC Reference accuracy ...
Страница 360: ...DM2571 User Manual 356 Dimensions All dimensions are shown in millimeters Optional GPIB card is installed ...
Страница 362: ......
Страница 363: ......
Страница 364: ......
Страница 366: ......
Страница 368: ......
Страница 369: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama223 8508JAPAN ...