STATus Report Commands
STATus:OPERation:CONDition?
Returns the total number of the Operation Condition register.
Return parameter: <NR1>, Ex: +4096
●A condition register continuously monitors the state of the instrument.
Condition register bits are updated in real time; they are neither latched nor
buffered.
This register is read-only; bits are not cleared when read.
STATus:OPERation:ENABle
Sets bits in the Operation Enable register.
Parameter: <NR1> (0 to 32767)
Example: STAT:OPER:ENAB 10
Sets the bit1 and bit3 in Operation Enable register, 10 = 2
1
+
2
3
.
●The selected bits are then reported to the Status Byte. An enable register
defines which bits in the event register will be reported to the Status Byte
register group. You can write to or read from an enable register.
●A STATus:PRESet clears all bits in the enable register.
●The *PSC command controls whether the enable register is cleared at power
on.
STATus:OPERation:ENABle?
Returns the total number of the Operation Enable register.
Return parameter: <NR1>, Ex: +256
STATus:OPERation[:EVENt]?
Returns the total number of the Operation Event register.
Return parameter: <NR1>, Ex: +786
●An event register is a read-only register that latches events from the
condition register. While an event bit is set, subsequent events corresponding
to that bit are ignored.
●Once a bit is set, it remains set until cleared by reading the event register or
by sending *CLS (clear status).
STATus:PRESet
Clears the Operation Enable register and Questionable Enable register.
Example: STAT:PRES
Содержание DM2571
Страница 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Страница 2: ......
Страница 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 76: ...DM2571 User Manual 72 ...
Страница 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Страница 176: ...DM2571 User Manual 172 ...
Страница 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Страница 185: ...DISPLAY SETTING 181 DISPLAY SETTING Digit 182 Display 184 Number 184 Bar Meter 185 Trend Chart 189 Histogram 198 ...
Страница 244: ...DM2571 User Manual 240 ...
Страница 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Страница 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Страница 348: ...DM2571 User Manual 344 DCV Ratio 5 Accuracy Specification DC Input accuracy DC Reference accuracy ...
Страница 360: ...DM2571 User Manual 356 Dimensions All dimensions are shown in millimeters Optional GPIB card is installed ...
Страница 362: ......
Страница 363: ......
Страница 364: ......
Страница 366: ......
Страница 368: ......
Страница 369: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama223 8508JAPAN ...