DISPLAY SETTING
199
F3 (Bins) key
to define bin
numbers
Backgro
und
Press the F3 (Bins) key to enter the Bins Setup
menu where user can customize the maximum
numbers of stripe-like bins for display.
Note: The available options for bin numbers will
vary in accordance with the defined refresh rate.
Faster the refresh rate, smaller the numbers of
bins available.
Display
The histogram is defined
with 20 bins display. The
central line divides the
left and right parts, each
of which contains 10
bins respectively.
The 10 bins setting
make the histogram
display much thicker in
its each bin compared to
the previous 20 bins
setting.
The max bin number varies by the refresh rate.
Check the table below for correlative parameters.
Refresh
Rate
5/s to
2.4k/s
4.8k/s
7.2k/s
10k/s
Max.
Bin
Number
400
200
100
20
F4 (HScale)
key to define
horizontal
scale
Backgro
und
Press the F4 (HScale) key to enter the HScale
Setup menu where Auto and Manual options are
available for selection.
Auto
Selecting “Auto” allows the
frequency of the
measuring counts to be symmetric with the
defined refresh rate. For example, setting 10k/s
results in the fastest frequency, while 1s leading
to the slowest frequency.
Содержание DM2571
Страница 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Страница 2: ......
Страница 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 76: ...DM2571 User Manual 72 ...
Страница 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Страница 176: ...DM2571 User Manual 172 ...
Страница 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Страница 185: ...DISPLAY SETTING 181 DISPLAY SETTING Digit 182 Display 184 Number 184 Bar Meter 185 Trend Chart 189 Histogram 198 ...
Страница 244: ...DM2571 User Manual 240 ...
Страница 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Страница 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Страница 348: ...DM2571 User Manual 344 DCV Ratio 5 Accuracy Specification DC Input accuracy DC Reference accuracy ...
Страница 360: ...DM2571 User Manual 356 Dimensions All dimensions are shown in millimeters Optional GPIB card is installed ...
Страница 362: ......
Страница 363: ......
Страница 364: ......
Страница 366: ......
Страница 368: ......
Страница 369: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama223 8508JAPAN ...