DM2571 User Manual
352
Operating Characteristics
Function
Gate Time
Digits
Frequency,
Period
1 s
6 ½
100 ms
5 ½
10 ms
4 ½
[1].
This specification will be available after 60 minutes of warm-up and
sine wave input, unless stated otherwise. This specification applies to 1s
gate time.
[2].
This specification is available when both sine wave and square wave
input ≥ 100 mV. For the input of 10 mV to 100 mV, the % of reading
error needs to be multiplied by 10 times.
[3].
The amplitude range is from 10% to 120% and is lower than 750
VAC.
[4].
The input ≥ 60 mV, for 300 k to 1 MHz, within 100mV range.
Temperature Characteristics
[1]
(Exclusive of probe errors)
RTD (Accuracy based on PT100):
(100 Ω platinum [PT100], D100, F100,PT385, PT3916, or user type)
Range
Resolution
1 Year
(TCAL ±5°C)
Temperature Coefficient
-200 °C to -100 °C
0.001 °C
0.09 °C
0.004 °C / °C
-100 °C to -20 °C
0.001 °C
0.08 °C
0.005 °C / °C
-20 °C to 20 °C
0.001 °C
0.06 °C
0.005 °C / °C
20 °C to 100 °C
0.001 °C
0.08 °C
0.005 °C / °C
100 °C to 300 °C
0.001 °C
0.12 °C
0.007 °C / °C
300 °C to 600 °C
0.001 °C
0.22 °C
0.009 °C / °C
Thermocouples
(Accuracy based on ITS-90):
Type Range
Resolution
1 Year
(TCAL±5 °C)*
Temperature Coefficient
E
-200 to +1000 °C
0.002 °C
0.2 °C
0.03 °C / °C
J
-210 to +1200
˚
C
0.002 °C
0.2 °C
0.03 °C / °C
T
-200 to +400 °C
0.002 °C
0.3 °C
0.04 °C / °C
K
-200 to +1372
˚
C
0.002 °C
0.3 °C
0.04 °C / °C
N
-200 to +1300
˚
C
0.003 °C
0.4 °C
0.05 °C / °C
R
-50 to +1768
˚
C
0.01 °C
1 °C
0.14 °C / °C
S
-50 to +1768
˚
C
0.01 °C
1 °C
0.14 °C / °C
B
+350 to +1820
˚
C
0.01 °C
1 °C
0.14 °C / °C
*Relative to simulated junction
Thermistor (2.2 kΩ, 5 kΩ, 10 kΩ or User Type)
Range
Resolution
1 Year (TCAL±5 °C) Temperature Coefficient
–80 ° to 150 °C
0.001 °C
0.1 °C
0.003 °C/ °C
Содержание DM2571
Страница 1: ...NF Corporation DIGITAL MULTIMETER DM2571 USER MANUAL ...
Страница 2: ......
Страница 3: ...DA00078064 002 DIGITAL MULTIMETER DM2571 USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 76: ...DM2571 User Manual 72 ...
Страница 137: ...SYSTEM FIRMWARE 133 SYSTEM FIRMWARE View System Info 134 Firmware Update 135 ...
Страница 176: ...DM2571 User Manual 172 ...
Страница 177: ...SCREENSHOT LOG 173 SCREENSHOT LOG Capture 174 Save Reading 177 ...
Страница 185: ...DISPLAY SETTING 181 DISPLAY SETTING Digit 182 Display 184 Number 184 Bar Meter 185 Trend Chart 189 Histogram 198 ...
Страница 244: ...DM2571 User Manual 240 ...
Страница 261: ...REMOTE CONTROL 257 SAV 328 SRE 328 SRE 328 STB 328 TRG 329 WAI 329 ...
Страница 334: ...DM2571 User Manual 330 Status system The diagram below is a description of the status system ...
Страница 348: ...DM2571 User Manual 344 DCV Ratio 5 Accuracy Specification DC Input accuracy DC Reference accuracy ...
Страница 360: ...DM2571 User Manual 356 Dimensions All dimensions are shown in millimeters Optional GPIB card is installed ...
Страница 362: ......
Страница 363: ......
Страница 364: ......
Страница 366: ......
Страница 368: ......
Страница 369: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama223 8508JAPAN ...