NPB-40 Operation
NPB-40
61
Note
: The %SpO
2
upper alarm limit range is 21% to
100%. The lower value of the %SpO
2
upper alarm
limit is limited to the %SpO
2
low alarm limit. The
%SpO
2
upper alarm limit cannot be set equal to or
lower than the %SpO
2
low alarm limit.
4. Press the
Up Arrow
or
Down Arrow
buttons to
increase or decrease the shown limit parameter.
5. Press the
Menu
button to set the limit value. The
BPM lower alarm limit is shown.
Note
: The pulse rate low alarm limit range is 30 to 249.
The upper value of the pulse rate low alarm limit is
limited one number lower than the pulse rate upper
alarm limit. The pulse rate low alarm limit cannot
be set equal to or higher than the pulse rate upper
alarm limit.
Содержание Oximax NPB-40
Страница 1: ......
Страница 8: ...Blank Page...
Страница 14: ...Blank Page...
Страница 32: ...Blank Page...
Страница 38: ...Blank Page...
Страница 80: ...Blank Page...
Страница 82: ...NPB 40 Operation 74...
Страница 98: ...OXIMAX Sensors and Accessories 90 Blank Page...
Страница 104: ...Blank Page...
Страница 105: ...Performance Considerations NPB 40 97...
Страница 106: ...Performance Considerations 98...
Страница 116: ...Blank Page...
Страница 120: ...Blank Page...
Страница 125: ...Menu Structure NPB 40 117 31 November 29 February on a non leap year...
Страница 126: ...Blank Page...
Страница 132: ...Blank Page...
Страница 151: ...Specifications NPB 40 143...
Страница 152: ...Specifications 144...
Страница 158: ...Index 152 performance 91 W warning 1 water resistant jacket 87 wavelength 132 wavelength sensor 132 Y year 116...
Страница 159: ...Index NPB 40 153...
Страница 160: ......