Troubleshooting
NPB-40
101
Table 5 lists the error codes and corrective actions. When
an error occurs, the unit will stop monitoring, remove all
information from the display and display message “Er
XXX,” and sound a low priority alarm. Turn the NPB-40
off, wait 10 seconds, then turn the NPB-40 on. If the
error appears, follow the action(s) listed in Table 5.
Table 5: Error Codes
Error
Code
Action
10
Check/replace sensor/extension cable.
11
1 —
Replace batteries.
2 —
Notify service personnel.
17
Check/replace sensor/extension cable.
19
Check/replace sensor/extension cable.
273
1 —
Restart the NPB-40.
2 — Set the time and date.
3 —
Notify service personnel.
274
Return NPB-40 for reprogramming.
275
Check/replace sensor/extension cable.
276
Replace with Oximax sensor.
277
Check/replace sensor/extension cable.
280
Check/replace sensor/extension cable.
282
Check/replace sensor/extension cable.
521
Replace batteries.
522
Replace batteries.
Содержание Oximax NPB-40
Страница 1: ......
Страница 8: ...Blank Page...
Страница 14: ...Blank Page...
Страница 32: ...Blank Page...
Страница 38: ...Blank Page...
Страница 80: ...Blank Page...
Страница 82: ...NPB 40 Operation 74...
Страница 98: ...OXIMAX Sensors and Accessories 90 Blank Page...
Страница 104: ...Blank Page...
Страница 105: ...Performance Considerations NPB 40 97...
Страница 106: ...Performance Considerations 98...
Страница 116: ...Blank Page...
Страница 120: ...Blank Page...
Страница 125: ...Menu Structure NPB 40 117 31 November 29 February on a non leap year...
Страница 126: ...Blank Page...
Страница 132: ...Blank Page...
Страница 151: ...Specifications NPB 40 143...
Страница 152: ...Specifications 144...
Страница 158: ...Index 152 performance 91 W warning 1 water resistant jacket 87 wavelength 132 wavelength sensor 132 Y year 116...
Страница 159: ...Index NPB 40 153...
Страница 160: ......