NPB-40 Operation
NPB-40
63
8. Press the
Up Arrow
or
Down Arrow
buttons to
increase or decrease the shown limit parameter.
9. Press the
Menu
button to set the alarm limit value.
10. Press the
Menu
button four times to return to normal
operation.
Note
: Limit changes will only be in effect as long as the
NPB-40 remains turned on. When the
NPB-40
is
turned off, the default limits will be restored in the
NPB-40
. When the
NPB-40
is turned on, the default
limits will be in effect.
Alarm Limit Changed Indicator
Alarm limits that have been changed from the default
values are identified by a decimal point (.) after the
shown value (%SpO
2
or BPM).