Description of Controls, Indicators, and Symbols
18
Selected display elements (icons, numerals, etc.), may
flash. There are three flash rates. See Table 1.
%SpO2
display. Shows the oxygen saturation level of
functional hemoglobin. Displays two dashes throughout
Sensor Disconnected and Sensor Off Patient conditions.
The display flashes the SpO
2
value when the SpO
2
is
outside the alarm limits. If alarm limits have been
changed from their power
-
on defaults, a decimal point (.)
is shown after the SpO
2
value (98.).
Pulse Amplitude
indicator (blip bar). The
Pulse
Amplitude
indicator indicates the dynamic pulse
amplitude and rate. As the detected pulse becomes
stronger, more bars light with each pulse.
In the parameter-set menu, the
Pulse Amplitude
indicator displays with the number of shown elements
incremenated/decremented to reflect changes in the
volume level for alarms and
Pulse Amplitude
indicator
(blip bar).
Table 1: Flash Rates
Priority
Hertz
Duty Cycle
High
1.4 Hz to 2.8 Hz
20% to 60%
Medium
0.4 Hz to 0.8 Hz
20% to 60%
Low
N/A
Constant on
Содержание Oximax NPB-40
Страница 1: ......
Страница 8: ...Blank Page...
Страница 14: ...Blank Page...
Страница 32: ...Blank Page...
Страница 38: ...Blank Page...
Страница 80: ...Blank Page...
Страница 82: ...NPB 40 Operation 74...
Страница 98: ...OXIMAX Sensors and Accessories 90 Blank Page...
Страница 104: ...Blank Page...
Страница 105: ...Performance Considerations NPB 40 97...
Страница 106: ...Performance Considerations 98...
Страница 116: ...Blank Page...
Страница 120: ...Blank Page...
Страница 125: ...Menu Structure NPB 40 117 31 November 29 February on a non leap year...
Страница 126: ...Blank Page...
Страница 132: ...Blank Page...
Страница 151: ...Specifications NPB 40 143...
Страница 152: ...Specifications 144...
Страница 158: ...Index 152 performance 91 W warning 1 water resistant jacket 87 wavelength 132 wavelength sensor 132 Y year 116...
Страница 159: ...Index NPB 40 153...
Страница 160: ......