Set up the NPB-40
28
Caution: Use only Nellcor-approved
O
XI
M
AX
sensors
and extension cables.
Note
: Physiological conditions, medical procedures, or
external agents that may interfere with the
NPB-40’s ability to detect and display
measurements include:
•
dysfunctional hemoglobin
•
arterial dyes
•
low perfusion
•
dark pigment
•
externally applied coloring agents, such as nail
polish, dye, or pigmented cream
Inaccurate measurements can be caused by:
•
incorrect application of the
O
XI
M
AX
sensor
•
placement of the
O
XI
M
AX
sensor on an extremity
with a blood pressure cuff, arterial catheter, or
intravascular line
•
ambient light
•
prolonged and/or excessive patient movement
•
intravascular dyes or externally applied coloring,
such as nail polish or pigmented cream
•
failure to cover the
O
XI
M
AX
sensor site with opaque
material in high ambient light conditions
Содержание Oximax NPB-40
Страница 1: ......
Страница 8: ...Blank Page...
Страница 14: ...Blank Page...
Страница 32: ...Blank Page...
Страница 38: ...Blank Page...
Страница 80: ...Blank Page...
Страница 82: ...NPB 40 Operation 74...
Страница 98: ...OXIMAX Sensors and Accessories 90 Blank Page...
Страница 104: ...Blank Page...
Страница 105: ...Performance Considerations NPB 40 97...
Страница 106: ...Performance Considerations 98...
Страница 116: ...Blank Page...
Страница 120: ...Blank Page...
Страница 125: ...Menu Structure NPB 40 117 31 November 29 February on a non leap year...
Страница 126: ...Blank Page...
Страница 132: ...Blank Page...
Страница 151: ...Specifications NPB 40 143...
Страница 152: ...Specifications 144...
Страница 158: ...Index 152 performance 91 W warning 1 water resistant jacket 87 wavelength 132 wavelength sensor 132 Y year 116...
Страница 159: ...Index NPB 40 153...
Страница 160: ......