Troubleshooting
NPB-40
103
1. There is no response to the
Power
button.
•
Ensure that the
Power
button is fully depressed.
•
The batteries may be missing, discharged, or
oriented incorrectly. Install new batteries. See
2. One or more display segments or indicators do
not light during the power-on self-test.
•
Do not use the NPB-40; contact qualified service
personnel or your local Nellcor representative.
3. The
Pulse Search
indicator is displayed for more
than 10 seconds.
•
Check the sensor directions for use to determine if
an appropriate sensor is being used and if it is
applied properly. Check sensor and extension cable
connections. Test the sensor on another subject.
Try another sensor or extension cable.
•
Perfusion may be too low for the NPB-40 to track
the pulse. Check the patient. Test the NPB-40 on
yourself. Change the sensor site. Try another
sensor.
•
Excessive patient motion may be preventing the
NPB-40 from tracking the pulse. Keep the patient
still, if possible. Verify that the sensor is securely
applied, and replace it if necessary. Change the
sensor site.
Содержание Oximax NPB-40
Страница 1: ......
Страница 8: ...Blank Page...
Страница 14: ...Blank Page...
Страница 32: ...Blank Page...
Страница 38: ...Blank Page...
Страница 80: ...Blank Page...
Страница 82: ...NPB 40 Operation 74...
Страница 98: ...OXIMAX Sensors and Accessories 90 Blank Page...
Страница 104: ...Blank Page...
Страница 105: ...Performance Considerations NPB 40 97...
Страница 106: ...Performance Considerations 98...
Страница 116: ...Blank Page...
Страница 120: ...Blank Page...
Страница 125: ...Menu Structure NPB 40 117 31 November 29 February on a non leap year...
Страница 126: ...Blank Page...
Страница 132: ...Blank Page...
Страница 151: ...Specifications NPB 40 143...
Страница 152: ...Specifications 144...
Страница 158: ...Index 152 performance 91 W warning 1 water resistant jacket 87 wavelength 132 wavelength sensor 132 Y year 116...
Страница 159: ...Index NPB 40 153...
Страница 160: ......