NPB-40 Operation
NPB-40
65
Procedure
With the NPB-40 in the normal mode:
1. Press the
Data
button to capture a snap-shot of the
NPB-40 data. The capture data display will be
shown.
Note:
The number (1) in the SpO
2
field is the
identification (ID) number of captured snap-shots.
Note:
When the
Data
button is pressed and there is no
empty event memory location available, the
NPB-40:
•
displays the last ID number assigned (50)
•
the
Data
icon continues to flash at the medium
priority rate
•
sounds an invalid key tone
•
returns to the normal operating mode immediately