Performance Considerations
NPB-40
95
High ambient light sources, such as surgical lights
(especially those with a xenon light source), bilirubin
lamps, fluorescent lights, infrared heating lamps, and
direct sunlight can interfere with the performance of an
SpO
2
O
XI
M
AX
sensor. To prevent interference from
ambient light, ensure that the
O
XI
M
AX
sensor is properly
applied, and cover the
O
XI
M
AX
sensor site with opaque
material.
WARNING: Failure to cover the
O
XI
M
AX
sensor site
with opaque material in high ambient light conditions
may result in inaccurate measurements.
If patient movement presents a problem, try one or more
of the following remedies to correct the problem.
•
verify that the
O
XI
M
AX
sensor is properly and
securely applied
•
move the
O
XI
M
AX
sensor to a less active site
•
use an adhesive
O
XI
M
AX
sensor that tolerates some
patient motion
•
use a new
O
XI
M
AX
sensor with fresh adhesive
backing
If poor perfusion affects performance, consider using the
MAX-R
O
XI
M
AX
sensor or the MAX-FAST
O
XI
M
AX
sensor. The MAX-R
O
XI
M
AX
sensor obtains
measurements from the nasal septal anterior ethmoid
artery, an artery supplied by the internal carotid. These
O
XI
M
AX
sensors may obtain measurements when
peripheral perfusion is relatively poor.
Содержание Oximax NPB-40
Страница 1: ......
Страница 8: ...Blank Page...
Страница 14: ...Blank Page...
Страница 32: ...Blank Page...
Страница 38: ...Blank Page...
Страница 80: ...Blank Page...
Страница 82: ...NPB 40 Operation 74...
Страница 98: ...OXIMAX Sensors and Accessories 90 Blank Page...
Страница 104: ...Blank Page...
Страница 105: ...Performance Considerations NPB 40 97...
Страница 106: ...Performance Considerations 98...
Страница 116: ...Blank Page...
Страница 120: ...Blank Page...
Страница 125: ...Menu Structure NPB 40 117 31 November 29 February on a non leap year...
Страница 126: ...Blank Page...
Страница 132: ...Blank Page...
Страница 151: ...Specifications NPB 40 143...
Страница 152: ...Specifications 144...
Страница 158: ...Index 152 performance 91 W warning 1 water resistant jacket 87 wavelength 132 wavelength sensor 132 Y year 116...
Страница 159: ...Index NPB 40 153...
Страница 160: ......