Principles of Operation
122
Figure 11: Oxyhemoglobin Dissociation Curve
O
XI
M
AX
Technology
The NPB-40 is designed to use Nellcor
O
XI
M
AX
brand
sensors, which integrate the
O
XI
M
AX
technology. These
O
XI
M
AX
sensors can be identified by their deep lavender/
blue plug color. All
O
XI
M
AX
sensors contain a memory
chip carrying information about the
O
XI
M
AX
sensor,
which the oximeter needs for correct operation, including
the
O
XI
M
AX
sensor’s calibration data, model type,
troubleshooting codes, and error detection data.
When an
O
XI
M
AX
sensor is connected to the NPB-40, the
pulse oximeter will first reads the information in the
O
XI
M
AX
sensor memory chip, checks it to make sure that
there are no errors, and then loads the data to begin
Содержание Oximax NPB-40
Страница 1: ......
Страница 8: ...Blank Page...
Страница 14: ...Blank Page...
Страница 32: ...Blank Page...
Страница 38: ...Blank Page...
Страница 80: ...Blank Page...
Страница 82: ...NPB 40 Operation 74...
Страница 98: ...OXIMAX Sensors and Accessories 90 Blank Page...
Страница 104: ...Blank Page...
Страница 105: ...Performance Considerations NPB 40 97...
Страница 106: ...Performance Considerations 98...
Страница 116: ...Blank Page...
Страница 120: ...Blank Page...
Страница 125: ...Menu Structure NPB 40 117 31 November 29 February on a non leap year...
Страница 126: ...Blank Page...
Страница 132: ...Blank Page...
Страница 151: ...Specifications NPB 40 143...
Страница 152: ...Specifications 144...
Страница 158: ...Index 152 performance 91 W warning 1 water resistant jacket 87 wavelength 132 wavelength sensor 132 Y year 116...
Страница 159: ...Index NPB 40 153...
Страница 160: ......