Troubleshooting
104
•
The sensor may be too tight, there may be
excessive ambient light, or the sensor may be on an
extremity with a blood pressure cuff, arterial
catheter, or intravascular line. Reposition sensor,
as necessary.
•
Excessive environmental motion or
electromagnetic interference may be preventing the
NPB-40 from tracking the pulse. Remove the
source of interference or try to stabilize the
environment, or do both.
4. The
Pulse Search
indicator lights after successful
measurements have been made.
•
Check the patient.
•
Perfusion may be too low for the NPB-40 to track
the pulse. Test the NPB-40 on another subject.
Change the sensor site. Try another type of sensor.
•
Excessive patient motion may be preventing the
NPB-40 from tracking the pulse. Verify that the
sensor is securely applied and replace it if
necessary. Change the sensor site.
•
The sensor may be too tight, there may be
excessive ambient light, or the sensor may be on an
extremity with a blood pressure cuff, arterial
catheter, or intravascular line. Reposition sensor,
as necessary.
Содержание Oximax NPB-40
Страница 1: ......
Страница 8: ...Blank Page...
Страница 14: ...Blank Page...
Страница 32: ...Blank Page...
Страница 38: ...Blank Page...
Страница 80: ...Blank Page...
Страница 82: ...NPB 40 Operation 74...
Страница 98: ...OXIMAX Sensors and Accessories 90 Blank Page...
Страница 104: ...Blank Page...
Страница 105: ...Performance Considerations NPB 40 97...
Страница 106: ...Performance Considerations 98...
Страница 116: ...Blank Page...
Страница 120: ...Blank Page...
Страница 125: ...Menu Structure NPB 40 117 31 November 29 February on a non leap year...
Страница 126: ...Blank Page...
Страница 132: ...Blank Page...
Страница 151: ...Specifications NPB 40 143...
Страница 152: ...Specifications 144...
Страница 158: ...Index 152 performance 91 W warning 1 water resistant jacket 87 wavelength 132 wavelength sensor 132 Y year 116...
Страница 159: ...Index NPB 40 153...
Страница 160: ......