Set up the NPB-40
NPB-40
29
Loss-of-pulse signal can occur for the following reasons:
•
the
O
XI
M
AX
sensor is applied too tightly
•
a blood pressure cuff is inflated on the same
extremity as the one with the
O
XI
M
AX
sensor
attached
•
there is arterial occlusion proximal to the
O
XI
M
AX
sensor
•
poor peripheral perfusion
Connect a Nellcor
O
XI
M
AX
SpO
2
sensor to the NPB-40
SpO
2
sensor port.
Note: An extension cable may be used to provide more
distance between the NPB-40 and the sensor. Use
the DEC-4 extension cable available from Nellcor:
•
DEC-4, 4 feet (1.2 m)
1
—
SpO
2
O
XI
M
AX
Sensor Port