Performance Considerations
N-600x Operator’s Manual
137
WARNING: Failure to cover the O
XI
M
AX
sensor site with opaque
material in high ambient light conditions may result in inaccurate
measurements.
If patient activity presents a problem, try one or more of the following
remedies to correct the problem.
•
verify the
O
XI
M
AX
sensor is properly and securely applied
•
move the
O
XI
M
AX
sensor to a less active site
•
use an adhesive
O
XI
M
AX
sensor that improves patient skin
contact
•
use a new
O
XI
M
AX
sensor with fresh adhesive backing
•
keep the patient still, if possible
If poor perfusion affects performance, consider using the MAX-R
O
XI
M
AX
sensor; it obtains measurements from the nasal septal anterior
ethmoid artery, an artery supplied by the internal carotid. This
O
XI
M
AX
sensor may obtain measurements when peripheral perfusion
is relatively poor.
Содержание OXIMAX N-600X Series
Страница 1: ......
Страница 8: ...Contents vi...
Страница 12: ...Safety Information 4 N 600x Operator s Manual...
Страница 80: ...Using the Monitor 72 N 600x Operator s Manual...
Страница 92: ...Using Monitor Trend Data 84 N 600x Operator s Manual...
Страница 106: ...Using OXIMAX Sensor Event Records 98 N 600x Operator s Manual...
Страница 109: ...Printing Monitor Trend Data N 600x Operator s Manual 101 12 Press the PRINT softkey ASCII printout GRAPH printout...
Страница 118: ...Printing Monitor Trend Data 110 N 600x Operator s Manual...
Страница 128: ...Using the Data Port 120 N 600x Operator s Manual...
Страница 140: ...OXIMAX Sensors and Accessories 132 N 600x Operator s Manual...
Страница 146: ...Performance Considerations 138 N 600x Operator s Manual...
Страница 168: ...Troubleshooting 160 N 600x Operator s Manual...
Страница 174: ...Using SatSeconds 166 N 600x Operator s Manual...
Страница 178: ...Factory Default Settings 170 N 600x Operator s Manual...
Страница 204: ...Specifications 196 N 600x Operator s Manual...
Страница 210: ...Index 202 N 600x Operator s Manual...
Страница 211: ......