Using OXIMAX Sensor Event Records
N-600x Operator’s Manual
85
U s i n g O
X I
M
A X
S e n s o r E v e n t R e c o r d s
Overview
WARNING: The sensor extrapolates from the date and time
provided by the N
-
600x when recording the sensor event record
to the sensor. The accuracy of the date/time is the responsibility of
the N
-
600x. It is recommended that the N
-
600x user set the time/
date to the correct value before a sensor event record
-
enabled
sensor is connected, and that this date/time not be changed while
the sensor remains connected. Since a sensor with sensor event
record data can be transported from one monitor to another,
having discrepancies in the date/time between monitors and the
sensor event record data will affect the order the sensor event
record data appears. To eliminate this possible problem, all
monitors within an institution should be set to the same time.
The adhesive
O
XI
M
AX
sensors are capable of storing patient alarm
event data. A sensor event record allows alarm event history to travel
with the patient on the sensor’s memory chip for quick assessment at
every point of care where
O
XI
M
AX
monitors are used.
Patient (alarm event) data is stored on the memory chip of adhesive
O
XI
M
AX
sensors (single-patient-use
O
XI
M
AX
sensors only). The alarm
event data is stored (recorded) with the limit/threshold settings that
were active at the time of the event on the recording monitor. These
events can be viewed on the next
O
XI
M
AX
sensor monitor when the
patient moves to a new point of care.
An event occurs when the %SpO
2
value exceeds either the upper or
lower alarm limit for at least 15 seconds. Alarm events are grouped
and recorded to the memory chip every 5 minutes. The maximum
number of events that can be stored in an
O
XI
M
AX
sensor is typically
100.
Содержание OXIMAX N-600X Series
Страница 1: ......
Страница 8: ...Contents vi...
Страница 12: ...Safety Information 4 N 600x Operator s Manual...
Страница 80: ...Using the Monitor 72 N 600x Operator s Manual...
Страница 92: ...Using Monitor Trend Data 84 N 600x Operator s Manual...
Страница 106: ...Using OXIMAX Sensor Event Records 98 N 600x Operator s Manual...
Страница 109: ...Printing Monitor Trend Data N 600x Operator s Manual 101 12 Press the PRINT softkey ASCII printout GRAPH printout...
Страница 118: ...Printing Monitor Trend Data 110 N 600x Operator s Manual...
Страница 128: ...Using the Data Port 120 N 600x Operator s Manual...
Страница 140: ...OXIMAX Sensors and Accessories 132 N 600x Operator s Manual...
Страница 146: ...Performance Considerations 138 N 600x Operator s Manual...
Страница 168: ...Troubleshooting 160 N 600x Operator s Manual...
Страница 174: ...Using SatSeconds 166 N 600x Operator s Manual...
Страница 178: ...Factory Default Settings 170 N 600x Operator s Manual...
Страница 204: ...Specifications 196 N 600x Operator s Manual...
Страница 210: ...Index 202 N 600x Operator s Manual...
Страница 211: ......