Troubleshooting
N-600x Operator’s Manual
147
575
TRENDS
LOST
Monitor trends are corrupted and will
be cleared.
Turn off the monitor, and then back on
again.
701-716,
720-724,
732-740,
576-582
POWER
SUPPLY
FAILURE
The monitor power supply has detected
an error. The monitor will shutdown
after 10 seconds.
Verify the monitor is being operated
within specified environmental
conditions. Notify a qualified service
technician.
717, 718
BATTERY
FAILURE
The monitor has detected a battery open
or short condition. The monitor will
shutdown after 10 seconds.
Battery should be replaced. Notify a
qualified service technician.
725-728,
730
REPLACE
BATTERY
The battery is not charging properly.
The monitor will shutdown after 10
seconds.
Battery should be replaced. Notify a
qualified service technician.
Table 12: Error Codes
Error
Code
Error
Message
Action
Содержание OXIMAX N-600X Series
Страница 1: ......
Страница 8: ...Contents vi...
Страница 12: ...Safety Information 4 N 600x Operator s Manual...
Страница 80: ...Using the Monitor 72 N 600x Operator s Manual...
Страница 92: ...Using Monitor Trend Data 84 N 600x Operator s Manual...
Страница 106: ...Using OXIMAX Sensor Event Records 98 N 600x Operator s Manual...
Страница 109: ...Printing Monitor Trend Data N 600x Operator s Manual 101 12 Press the PRINT softkey ASCII printout GRAPH printout...
Страница 118: ...Printing Monitor Trend Data 110 N 600x Operator s Manual...
Страница 128: ...Using the Data Port 120 N 600x Operator s Manual...
Страница 140: ...OXIMAX Sensors and Accessories 132 N 600x Operator s Manual...
Страница 146: ...Performance Considerations 138 N 600x Operator s Manual...
Страница 168: ...Troubleshooting 160 N 600x Operator s Manual...
Страница 174: ...Using SatSeconds 166 N 600x Operator s Manual...
Страница 178: ...Factory Default Settings 170 N 600x Operator s Manual...
Страница 204: ...Specifications 196 N 600x Operator s Manual...
Страница 210: ...Index 202 N 600x Operator s Manual...
Страница 211: ......