Principles of Operation
174
N-600x Operator’s Manual
Measured versus Calculated Saturation
When saturation is calculated from a blood gas partial pressure of
oxygen (PO
2
), the calculated value may differ from the SpO
2
measurement of a pulse oximeter. This usually occurs because the
calculated saturation was not properly corrected for the effects of
variables that shift the relationship between PO
2
and pH, temperature,
the partial pressure of carbon dioxide (PCO
2
), 2,3-DPG, and fetal
hemoglobin. See Figure 12.
Figure 12: Oxyhemoglobin Dissociation Curve
Содержание OXIMAX N-600X Series
Страница 1: ......
Страница 8: ...Contents vi...
Страница 12: ...Safety Information 4 N 600x Operator s Manual...
Страница 80: ...Using the Monitor 72 N 600x Operator s Manual...
Страница 92: ...Using Monitor Trend Data 84 N 600x Operator s Manual...
Страница 106: ...Using OXIMAX Sensor Event Records 98 N 600x Operator s Manual...
Страница 109: ...Printing Monitor Trend Data N 600x Operator s Manual 101 12 Press the PRINT softkey ASCII printout GRAPH printout...
Страница 118: ...Printing Monitor Trend Data 110 N 600x Operator s Manual...
Страница 128: ...Using the Data Port 120 N 600x Operator s Manual...
Страница 140: ...OXIMAX Sensors and Accessories 132 N 600x Operator s Manual...
Страница 146: ...Performance Considerations 138 N 600x Operator s Manual...
Страница 168: ...Troubleshooting 160 N 600x Operator s Manual...
Страница 174: ...Using SatSeconds 166 N 600x Operator s Manual...
Страница 178: ...Factory Default Settings 170 N 600x Operator s Manual...
Страница 204: ...Specifications 196 N 600x Operator s Manual...
Страница 210: ...Index 202 N 600x Operator s Manual...
Страница 211: ......