Using SatSeconds
N-600x Operator’s Manual
163
U s i n g S a t S e c o n d s
Overview
With traditional alarm management, upper and lower alarm limits are
set for monitoring oxygen saturation. During monitoring, as soon as
an alarm limit is violated by as little as one percentage point, an
audible alarm immediately sounds. When the %SpO
2
level fluctuates
near an alarm limit, the alarm sounds each time the limit is violated.
Such frequent alarms can be distracting.
The N-600x pulse oximeter utilizes Nellcor
SatSeconds
alarm
management technique. With the
SatSeconds
technique, upper and
lower alarm limits are set in the same way as with traditional alarm
management. The clinician also sets a
SatSeconds
limit that allows
the monitoring of %SpO
2
below the selected low alarm limit for a
period of time before an audible alarm sounds.
The
SatSeconds
limit controls the time that the %SpO
2
level may fall
outside the alarm before an audible alarm sounds.
The method of calculation is as follows:
The number of percentage points that the %SpO
2
falls outside of the
alarm limit is multiplied by the number of seconds that the %SpO
2
level remains outside that limit. This can be stated as an equation:
Points x Seconds =
SatSeconds
Where:
Points = %SpO
2
percentage points outside of the limit
Seconds = number of seconds the %SpO
2
remains at that point
outside of the limit
Содержание OXIMAX N-600X Series
Страница 1: ......
Страница 8: ...Contents vi...
Страница 12: ...Safety Information 4 N 600x Operator s Manual...
Страница 80: ...Using the Monitor 72 N 600x Operator s Manual...
Страница 92: ...Using Monitor Trend Data 84 N 600x Operator s Manual...
Страница 106: ...Using OXIMAX Sensor Event Records 98 N 600x Operator s Manual...
Страница 109: ...Printing Monitor Trend Data N 600x Operator s Manual 101 12 Press the PRINT softkey ASCII printout GRAPH printout...
Страница 118: ...Printing Monitor Trend Data 110 N 600x Operator s Manual...
Страница 128: ...Using the Data Port 120 N 600x Operator s Manual...
Страница 140: ...OXIMAX Sensors and Accessories 132 N 600x Operator s Manual...
Страница 146: ...Performance Considerations 138 N 600x Operator s Manual...
Страница 168: ...Troubleshooting 160 N 600x Operator s Manual...
Страница 174: ...Using SatSeconds 166 N 600x Operator s Manual...
Страница 178: ...Factory Default Settings 170 N 600x Operator s Manual...
Страница 204: ...Specifications 196 N 600x Operator s Manual...
Страница 210: ...Index 202 N 600x Operator s Manual...
Страница 211: ......