Using the Data Port
112
N-600x Operator’s Manual
Connecting to the Data Port
The N-600x data port may be connected to a serial printer or PC by
using a cable terminated with:
•
an AMP connector (AMP part number 747538-1),
•
ferrule (AMP part number 1-747579-2), and
•
compatible pins (AMP part number 66570-2).
The cable should not exceed 25 feet (7.6 meters) in length. The
external ITE (Information Technology Equipment) device must be
certified to UL-1950 or IEC-60950. The cable used must have a
braided shield that provides 100% coverage, such as a Belden cable
(Belden part number 9609) or equivalent. The shield must have a
360-degree connection to the metal shell on the N-600x's DB-15
connector and to the connector on the PC or serial printer.
Caution: Do not create sharp bends in the cable, as this may tear
or break the shielding.
No hardware flow control is used. However, in ASCII mode, XON/
XOFF flow control is supported.
Data Port Pinouts
The pinouts for the data port are listed in Table 8.
Table 8: Data Port Pinouts
Pin
Signal Name
1
RXD+ (RS
-
422 [+] input)
2
RXD_232 (RS
-
232 input)
3
TXD_(RS
-
232 output)
4
TXD+ (RS
-
422 [+] output)
Содержание OXIMAX N-600X Series
Страница 1: ......
Страница 8: ...Contents vi...
Страница 12: ...Safety Information 4 N 600x Operator s Manual...
Страница 80: ...Using the Monitor 72 N 600x Operator s Manual...
Страница 92: ...Using Monitor Trend Data 84 N 600x Operator s Manual...
Страница 106: ...Using OXIMAX Sensor Event Records 98 N 600x Operator s Manual...
Страница 109: ...Printing Monitor Trend Data N 600x Operator s Manual 101 12 Press the PRINT softkey ASCII printout GRAPH printout...
Страница 118: ...Printing Monitor Trend Data 110 N 600x Operator s Manual...
Страница 128: ...Using the Data Port 120 N 600x Operator s Manual...
Страница 140: ...OXIMAX Sensors and Accessories 132 N 600x Operator s Manual...
Страница 146: ...Performance Considerations 138 N 600x Operator s Manual...
Страница 168: ...Troubleshooting 160 N 600x Operator s Manual...
Страница 174: ...Using SatSeconds 166 N 600x Operator s Manual...
Страница 178: ...Factory Default Settings 170 N 600x Operator s Manual...
Страница 204: ...Specifications 196 N 600x Operator s Manual...
Страница 210: ...Index 202 N 600x Operator s Manual...
Страница 211: ......