Specifications
N-600x Operator’s Manual
185
Electric fast
transient/
burst
IEC
61000
-
4
-
4
±2 kV for
power supply
lines
±1 kV for
input/output
lines
±2 kV for
power supply
lines
±1 kV for
input/output
lines
Mains power quality
should be that of a
typical commercial
and/or hospital
environment.
Surge
IEC
61000
-
4
-
5
±1 kV
differential
mode
±2 kV
common mode
±1 kV
differential
mode
±2 kV
common mode
Mains power quality
should be that of a
typical commercial
and/or hospital
environment.
Table 18: Electromagnetic Immunity
The N
-
600x is suitable for use in the specified electromagnetic
environment. The customer and/or user of the N
-
600x should assure that it
is used in an electromagnetic environment as described below.
Immunity
Test
IEC
60601
-
1
-
2
Test Level
Compliance
Level
Electromagnetic
Environment
Guidance
Note
: UT is the AC mains voltage prior to application of the test level.
Содержание OXIMAX N-600X Series
Страница 1: ......
Страница 8: ...Contents vi...
Страница 12: ...Safety Information 4 N 600x Operator s Manual...
Страница 80: ...Using the Monitor 72 N 600x Operator s Manual...
Страница 92: ...Using Monitor Trend Data 84 N 600x Operator s Manual...
Страница 106: ...Using OXIMAX Sensor Event Records 98 N 600x Operator s Manual...
Страница 109: ...Printing Monitor Trend Data N 600x Operator s Manual 101 12 Press the PRINT softkey ASCII printout GRAPH printout...
Страница 118: ...Printing Monitor Trend Data 110 N 600x Operator s Manual...
Страница 128: ...Using the Data Port 120 N 600x Operator s Manual...
Страница 140: ...OXIMAX Sensors and Accessories 132 N 600x Operator s Manual...
Страница 146: ...Performance Considerations 138 N 600x Operator s Manual...
Страница 168: ...Troubleshooting 160 N 600x Operator s Manual...
Страница 174: ...Using SatSeconds 166 N 600x Operator s Manual...
Страница 178: ...Factory Default Settings 170 N 600x Operator s Manual...
Страница 204: ...Specifications 196 N 600x Operator s Manual...
Страница 210: ...Index 202 N 600x Operator s Manual...
Страница 211: ......