National Instruments Corporation
69
NI 5402/5406 Calibration Procedure
4.
Set the pre-amplifier attenuation by calling the niFgen Property Node
and selecting
Calibration»Pre-Amplifier Attenuation
.
5.
Take the following voltage measurements at the NI 5402/5406 CH 0
front panel connector into a high-impedance load:
a.
Set the gain DAC value by calling the niFgen Property Node and
selecting
Calibration»Gain DAC Value
.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Set
AttributeViReal64
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
attributeID
:
NIFGEN_
ATTR_PRE_AMPLIFIER_
ATTENUATION
value
: The
Pre-Amplifier
Attenuation
value for the
current iteration from
Table 8
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Set
AttributeViInt32
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
attributeID
:
NIFGEN_
ATTR_GAIN_DAC_VALUE
value
:
1,500