NI 5402/5406 Calibration Procedure
32
ni.com
14
FFFF8000
4
0
–2
±0.005
±0.022
15
7FFF
3
0
1.5
±0.004
±0.017
16
FFFF8000
3
0
–1.5
±0.004
±0.017
17
7FFF
2.8
0
1.4
±0.0038
±0.016
18
FFFF8000
2.8
0
–1.4
±0.0038
±0.016
19
7FFF
2
0
1
±0.003
±0.012
20
FFFF8000
2
0
–1
±0.003
±0.012
21
7FFF
1.4
0
0.7
±0.0024
±0.009
22
FFFF8000
1.4
0
–0.7
±0.0024
±0.009
23
7FFF
1
0
0.5
±0.002
±0.007
24
FFFF8000
1
0
–0.5
±0.002
±0.007
25
7FFF
0.7
0
0.35
±0.0017
±0.0055
26
FFFF8000
0.7
0
–0.35
±0.0017
±0.0055
27
7FFF
0.5
0
0.25
±0.0015
±0.0045
28
FFFF8000
0.5
0
–0.25
±0.0015
±0.0045
29
7FFF
0.3
0
0.15
±0.0013
±0.0035
30
FFFF8000
0.3
0
–0.15
±0.0013
±0.0035
31
7FFF
0.2
0
0.1
±0.0012
±0.003
32
FFFF8000
0.2
0
–0.1
±0.0012
±0.003
33
7FFF
0.14
0
0.07
±0.00114
±0.0027
34
FFFF8000
0.14
0
–0.07
±0.00114
±0.0027
35
7FFF
0.1
0
0.05
±0.0011
±0.0025
36
FFFF8000
0.1
0
–0.05
±0.0011
±0.0025
37
7FFF
0.07
0
0.035
±0.00107
±0.00235
38
FFFF8000
0.07
0
–0.035
±0.00107
±0.00235
39
7FFF
0.05
0
0.025
±0.00105
±0.00225
40
FFFF8000
0.05
0
–0.025
±0.00105
±0.00225
41
7FFF
0.04
0
0.02
±0.00104
±0.0022
42
FFFF8000
0.04
0
–0.02
±0.00104
±0.0022
Table 6.
Values for Verifying Square Wave Gain and Offset Accuracy (Continued)
Iteration
Analog Static
Value
Amplitude
(V
pp
)
DC Offset
(V)
Ideal DC
Output
(V)
Calibration
Test Limits
(V)
Published
Specification
(V)