NI 5402/5406 Calibration Procedure
88
ni.com
13. Configure the flatness by calling the niFgen Cal Adjust Flatness VI.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Cal
AdjustFlatness
using
the following parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
frequenciesArray
: An
array of the frequencies
from Table 11
configuration
: The
Current Configuration
value for the current
iteration from Table 11
requestedAmplitude
AtCalibration
: The
requested amplitude
value for the current
configuration from
Table 10
measuredAmplitudes
:
An array of the
amplitudes measured by
the power meter during
the current iteration