National Instruments Corporation
27
NI 5402/5406 Calibration Procedure
10. Set the amplitude by calling the niFgen Property Node and selecting
Standard Function Output»Amplitude
.
11. Set the digital gain by calling the niFgen Property Node and selecting
Output Attributes»Digital Gain
.
LabVIEW Block Diagram
C/C++ Function Call
Call
niFgen_Set
AttributeViReal64
using the following
parameters:
vi
: The session handle
returned from
niFgen_init
.
channelName
:
"0"
attributeID
:
NIFGEN_
ATTR_FUNC_AMPLITUDE
value
: The
Amplitude
value for the current
iteration from Table 5
LabVIEW Block Diagram
C/C++ Function Call
Call
niFgen_Set
AttributeViReal64
using the following
parameters:
vi
: The session handle
returned from
niFgen_init
.
channelName
:
"0"
attributeID
:
NIFGEN_
ATTR_DIGITAL_GAIN
value
: The
Digital Gain
value for the current
iteration from Table 5