NI 5402/5406 Calibration Procedure
8.
Set the pre-amplifier attenuation by calling the niFgen Property Node
and selecting
Calibration»Pre-Amplifier Attenuation
.
9.
Set the amplitude by calling the niFgen Property Node and selecting
Standard Function Output»Amplitude
.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Set
AttributeViReal64
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
attributeID
:
NIFGEN_
ATTR_PRE_AMPLIFIER_
ATTENUATION
value
:
The
Pre-Amplifier
Attenuation
value for the
current iteration from
Table 10
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Set
AttributeViReal64
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
attributeID
:
NIFGEN_
ATTR_FUNC_AMPLITUDE
value
: The
Amplitude
value for the current
iteration from Table 10