NI 5402/5406 Calibration Procedure
10. Set the post-amplifier attenuation by calling the niFgen Property Node
and selecting
Calibration»Post-Amplifier Attenuation
.
11. Set the load impedance by calling the niFgen Property Node and
selecting
Output Attributes»Load Impedance
.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Set
AttributeViReal64
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
attributeID
:
NIFGEN_
ATTR_POST_AMPLIFIER
_ATTENUATION
value
:
0.00
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Set
AttributeViReal64
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
attributeID
:
NIFGEN_
ATTR_LOAD_IMPEDANCE
value
:
10,000,000,000