NI 5402/5406 Calibration Procedure
86
ni.com
10. Initiate waveform generation by calling the niFgen Initiate
Generation VI.
11. Generate a sine wave configured for the current iteration of Table 10 at
each of the frequencies listed in Table 11. These steps will be repeated
for each iteration in Table 10.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Initiate
Generation
using the following
parameter:
vi
: The session handle
returned from
niFgen_InitExtCal
.
Table 11.
Frequencies for Adjusting Sine Wave Flatness Correction
Frequencies for the NI 5402
(Hz)
Frequencies for the NI 5406
(Hz)
100,000
100,000
1,000,000
1,000,000
5,000,000
5,000,000
10,000,000
10,000,000
15,000,000
15,000,000
20,000,000
20,000,000
—
25,000,000
—
30,000,000
—
40,000,000