National Instruments Corporation
19
NI 5402/5406 Calibration Procedure
8.
Initiate waveform generation by calling the niFgen Initiate
Generation VI.
9.
Use the power meter to measure the output power level of the
NI 5402/5406. This value is the measured power of the 50 kHz sine
wave.
10. In this step you generate a sine wave configured for the current
iteration of Table 3 at each of the frequencies listed in Table 4.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Initiate
Generation
using the following
parameter:
vi
: The session handle
returned from
niFgen_init
.
Table 4.
Frequencies for Verifying Sine Wave Flatness Correction Accuracy
Frequencies for the NI 5402
(MHz)
Frequencies for the NI 5406
(MHz)
1.00000
1.00000
5.00000
5.00000
10.00000
10.00000
15.00000
15.00000
20.00000
20.00000
25.00000
25.00000
—
30.00000
—
40.00000