NI 5402/5406 Calibration Procedure
58
ni.com
8.
Set the analog filter state by calling the niFgen Property Node and
selecting
Output Attributes»Analog Filter Enabled
.
9.
Set the pre-amplifier attenuation by calling the niFgen Property Node
and selecting
Calibration»Pre-Amplifier Attenuation
.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_
SetAttribute
ViBoolean
using the
following parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
attributeID
:
NIFGEN_ATTR_ANALOG_
FILTER_ENABLED
value
:
VI_FALSE
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Set
AttributeViReal64
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
..
channelName
:
"0"
attributeID
:
NIFGEN_
ATTR_PRE_AMPLIFIER_
ATTENUATION
value
:
0.00