NI 5402/5406 Calibration Procedure
52
ni.com
30. Adjust the ADC calibration by calling the niFgen Cal Adjust Cal
ADC VI.
31. Disable the analog output by calling the niFgen Property Node and
selecting
Basic Operation»Output Enabled
.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Cal
AdjustCalADC
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
voltagesMeasured
Externally
: {
external
measurement 0
,
external
measurement 1
}
voltagesMeasured
WithCalADC
: {
cal
ADC measurement 0
,
cal ADC
measurement 1
}
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Set
AttributeViBoolean
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
attributeID
:
NIFGEN_
ATTR_OUTPUT_ENABLED
value
:
VI_FALSE