NI 5402/5406 Calibration Procedure
14
ni.com
Verifying the Sine Wave Flatness Correction
Accuracy
This test verifies the Sine Wave Flatness Correction Accuracy of the
NI 5402/5406. In this procedure, you will compare the signal generator’s
output power level during the generation of sine waves of various
frequencies to the output power level measured during the generation of a
50 kHz sine wave. The following figure shows the programming flow for
this test:
Figure 2.
Programming Flow for the Verification of the Sine Wave
Flatness Correction Accuracy
Initialize an NI-FGEN Session
Configure 50kHz Sine Wave
Set Analog Path, Amplitude,
Pre-amp Attenuation, and
Post-amp Attenuation of Waveform
Generate the Waveform
Measure Power Level
Change the Frequency
Measure Power Level
Repeat
for Each
Frequency
Abort Waveform Generation
Repeat
for Each
Configuration