NI 5402/5406 Calibration Procedure
26
ni.com
28
0
0.012
–0.269
–0.269
±0.000928
±0.003405
29
0
0.012
0.194
0.194
±0.000815
±0.00303
30
0
0.012
–0.194
–0.194
±0.000815
±0.00303
31
0
0.012
0.144
0.144
±0.00074
±0.00278
32
0
0.012
–0.144
–0.144
±0.00074
±0.00278
33
0
0.012
0.094
0.094
±0.000665
±0.00253
34
0
0.012
–0.094
–0.094
±0.000665
±0.00253
35
0
0.012
0.064
0.064
±0.00062
±0.00238
36
0
0.012
–0.064
–0.064
±0.00062
±0.00238
37
0
0.012
0.044
0.044
±0.00059
±0.00228
38
0
0.012
–0.044
–0.044
±0.00059
±0.00228
39
0
0.012
0.029
0.029
±0.000567
±0.002205
40
0
0.012
–0.029
–0.029
±0.000567
±0.002205
41
0
0.012
0.019
0.019
±0.000553
±0.002155
42
0
0.012
–0.019
–0.019
±0.000553
±0.002155
43
0
0.012
0.014
0.014
±0.000545
±0.00213
44
0
0.012
–0.014
–0.014
±0.000545
±0.00213
45
0
0.012
0.008
0.008
±0.000536
±0.0021
46
0
0.012
–0.008
–0.008
±0.000536
±0.0021
47
0
0.012
0.004
0.004
±0.00053
±0.00208
48
0
0.012
–0.004
–0.004
±0.00053
±0.00208
49
0
0.012
0.002
0.002
±0.000527
±0.00207
50
0
0.012
–0.002
–0.002
±0.000527
±0.00207
Table 5.
Values for Verifying DC Gain and Offset Accuracy (Continued)
Iteration
Digital Gain
Amplitude
(V
pp
)
DC Offset
(V)
Ideal DC
Output
(V)
Calibration
Test Limit
(V)
Published
Specification
(V)