NI 5402/5406 Calibration Procedure
56
ni.com
4.
Set the main DAC value by calling the niFgen Write Binary 16 Analog
Static Value VI.
5.
Set the analog path value by calling the niFgen Property Node and
selecting
Output Attributes»Analog Path
.
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Write
Binary16Analog
StaticValue
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
value
:
0
LabVIEW Block Diagram
C/C++
Function Call
Call
niFgen_Set
AttributeViInt32
using the following
parameters:
vi
: The session handle
returned from
niFgen_InitExtCal
.
channelName
:
"0"
attributeID
:
NIFGEN_ATTR_
ANALOG_PATH
value
:
NIFGEN_VAL_
FIXED_LOW_GAIN_
ANALOG_PATH