92
|
2017-07 MAN-C0034-06
USER MANUAL
nCounter
®
Dx Analysis System
2.
Select
Confirm
to continue the system shutdown.
FIGURE 7.3:
System shutdown messages displayed by the Prep Station (left) and the Digital Analyzer (right).
>>> The system will shut down.
3.
After the system shuts down (there is a delay of approximately 30 seconds), locate the power switch on the back of the instrument
and turn off the power.
4.
Wait an additional 30 seconds and then use the power switch to restore power to the instrument.
>>> The system will initialize and display the ‘Select Instrument Mode’ screen (see
Chapter 3 – Selecting the Instrument Mode
).
IMPORTANT:
To ensure proper communication between instruments after a power cycle, always restore power to the Digital
Analyzer first and allow the instrument to fully initialize before restoring power to the Prep Station.