c.
Touch
Next
.
>>> The ‘Tips and Foil Piercers’ screen will appear (
FIGURE 5.12
).
FIGURE 5.12:
The ‘Tips And Foil Piercers’ screen
10.
Instructions for the ‘Tips and Foil Piercers’ screen:
a.
Remove the metal tip carrier from the Prep Station deck by lifting straight up.
b.
Place the tips and the foil piercers into the carrier. Hold the plastic tip rack by the center tabs, hold the tip rack over the metal
holder, and slowly lower the tips into the metal holder. It is helpful to place the carrier at eye level to align the plastic tips
(
FIGURE 5.13
).
FIGURE 5.13:
Insert the tips and foil piercers into metal tip carrier
NOTE:
Each box contains two nested sets of tips; each side of the box contains one set of tips. When opening the box, hold a
hand firmly over the bottom of the box to avoid accidentally spilling the second set of tips.