72
|
2017-07 MAN-C0034-06
USER MANUAL
nCounter
®
Dx Analysis System
b.
Centrifuge the Reagent Plates at 2000 x g for 2 minutes to collect liquids at the bottom of the wells prior to loading the Reagent
Plates onto the Prep Station deck.
c.
While the Cartridges and Reagent Plates come to room temperature, continue setting up the Prep Station. Touch
Next
.
>>> The ‘Waste Receptacles’ screen will appear.
FIGURE 5.9:
The ‘Waste Receptacles’ screen
8.
Make sure the disposables from the previous run have been appropriately discarded. Press
Next
.
>>> The ‘Scan Reagent Plate’ screen will appear.
FIGURE 5.10:
The ‘Scan Reagent Plate’ screen