![NanoString Technologies nCounter Скачать руководство пользователя страница 89](http://html.mh-extra.com/html/nanostring-technologies/ncounter/ncounter_user-manual_705140089.webp)
2017-07 MAN-C0034-06
|
89
NanoString Technologies
®
USER MANUAL
5.
Confirm that a small blue bar appears at the bottom of the scanning Cartridge, indicating the scan has started. There will be a series
of rhythmic clicks as the images are collected.
6.
To add a Cartridge to a Digital Analyzer that is already counting, touch the
Pause
button on the ‘Counting Cartridges’ screen. It might
take a few minutes for the Digital Analyzer to reach a suitable stopping point (
FIGURE 6.7
). The door will unlock when it does. Place
the new Cartridge in an empty slot, or replace a Cartridge that has already been scanned. Touch the
Resume
button. The door should
re-lock and resume counting.
FIGURE 6.7:
The ‘Pause Counting Job’ screen
7.
When a Cartridge is finished scanning, an e-mail notification will be sent to the user and the processing screen will change to reflect
that the Cartridge has completed (
FIGURE 6.8
). Upon receiving the e-mail notification that the scan has completed, remove the
completed Cartridge. In case of an instrument error or reports not being available, store Cartridge in an opaque box (to protect from
light) at 4
o
C for up to one week. Contact
for assistance.
FIGURE 6.8:
The ‘Counting Cartridge’ screen with six completed Cartridges
8.
Using the link attached to the notification e-mail, open the web interface and download all the diagnostic reports associated with the
Run Set just completed as described in
Chapter 4 – Operating the Web Application.