2017-07 MAN-C0034-06
|
73
NanoString Technologies
®
USER MANUAL
9.
Instructions for the ‘Reagent Plate’ screen:
a.
The barcode for the Reagent Plates must be scanned before proceeding. Hold the barcode in front of the barcode reader. A red
beam will appear. Move the barcode in front of the red beam until it is recorded. When the barcode is successfully recorded, the
barcode number will appear in the box and the
Manual Entry
button will change to
Clear Entry
.
If you encounter any problems scanning the barcode, it may be entered manually. Touch
Manual Entry
; a numeric keypad will
appear. Touch the appropriate buttons to enter the number. If you make a mistake, touch
del
to backspace, or
clear
to start over.
Press
enter
when finished.
b.
Remove the clear plastic lids and place the Reagent Plates on the deck as indicated on the screen (
FIGURE 5.10
).
The deck has alignment pins that will allow the Reagent Plates to sit flat only if they are in the correct orientation. The Reagent
Plate should be oriented with the barcode facing the user (
FIGURE 5.11
). If the Reagent Plate is placed in the wrong direction, the
Prep Station will pause the protocol at the validation step until the user intervenes.
NOTE:
Only one Reagent Plate is required for runs performed using a 1-, 2-, 3-, or 4-test kit. For these kits, load the Reagent
Plate into the front position (nearest the user) on the Prep Station deck
NOTE:
The Prep Station will not accept barcodes for Reagent Plates that have expired. Make sure the Reagent Plates are used
prior to the expiration date.
FIGURE 5.11:
Use the positioning pins to ensure correct placement of the Reagent Plates.
IMPORTANT:
Do not remove the foil or pierce the wells on the Reagent Plates. The Prep Station pierces the aluminum foil
during processing.