2017-07 MAN-C0034-06
|
71
NanoString Technologies
®
USER MANUAL
6.
The barcode for the CodeSet must be scanned before proceeding. Hold the barcode of the CodeSet in front of the barcode reader. A
red beam will appear. Move the barcode in front of the red beam until it is read. When the barcode is successfully entered, the barcode
number will appear in the box (
FIGURE 5.6
).
NOTE:
The CodeSet barcode must match the CodeSet Kit Number entered when the Run Set was created using the web
application (
FIGURE 4.19
).
>>> The ‘Reagents And Cartridge’ screen will appear.
FIGURE 5.7:
The ‘Reagents And Cartridge’ screen
7.
Cartridges and Reagent Plate(s) (
FIGURE 5.8
) must be at room temperature prior to processing.
a.
Remove the nCounter Reagent Plates from storage at 4°C and the nCounter Cartridges from storage at -20°C. Allow them to
equilibrate to room temperature for 10–15 minutes.
FIGURE 5.8: A
sealed Sample Cartridge (left) and Dx Reagent Plate (right)
NOTE:
Only one Reagent Plate is required for runs performed using a 1-, 2-, 3-, or 4-test .
NOTE:
Do not open the Cartridge pouch until it has reached room temperature. This will prevent condensation from forming
on the Cartridge.